MR5320 Splash Proof Type Two-channel Card Logger
MR5320 is 2 ch type card logger of MR5300 and 6,000 data can be stored for a channel. Temperature sensor is only built in for channel 1. Data recalling and setting are also available on PC.
- Number of logging data is 6,000 data each for channels. Measurement interval can be selected between 10 to 50sec. and 1 to 60min.
- ±0.5°C (between -5to 50°C) of high accuracy
- Logging data can be stored even in dead battery.
- Dust tight and splash proof structure, so it is washable (IP64). Conform to the CE marking safety standard of Europe.
- Various external sensors: general type, waterproof type, and needle type etc. are available.
Models
Products | Models | Specifications |
---|---|---|
Logger | MR5320 | |
External Sensor | MR9301 | -40°C to 60°C |
MR9302 | 0°C to100°C | |
MR9303 | 50°C to150°C |
Specifications
Measuring Range | -40 to 60°C, 0 to 100°C, 50 to 150°C (Ranges are automatically switched by identifying a sensor to be used.) *Only channel 1 has built-in sensor, measuring range is -20 to 55°C |
---|---|
Accuracy Rating | ±0.5°C ±1digit (in a range of -5 to 50°C) *±1.0°C ±1digit of other than above. |
Input Point | 2 points |
Number of Logging Data | 6000 data for each channel |
Temperature Sensor | Built-in (channel 1 only)/External sensor switching *Various external sensors (sold separately): general type, waterproof tube, needle type and surface temperature measuring type etc. are available. |
Display Contents (Each channel) |
Current data=Month, day, hour, minute and measuring temperature Recalling data=Highest, and lowest temperatures in logging period, and average temperature in the logger Setting parameter=High and low limit alarm set values. |
Operation Available in the Logger | Display start-up (with auto-off function: Display goes out after one minute when no operation is performed.) Display content selection (clock setting, logging start/stop, keylock set/release) *Three function keys of DISP, SEL, and REC. |
Communication Function | Communication with an exclusive data reader (sold separately) through near-infrared radiation. Parameter setting, and recalling logger data and logging condition by exclusive analysis software. |
Setting in PC (Dedicated Analysis Software) |
Clock, logging start time/stop time (for logging interval unit: min.), measurement/logging interval (10 to 50 sec. or 1 to 60min.), high/low limit alarm temperatures, keylock of the logger, and logging method (automatic stop/endless) |
Power Supply | Alkaline battery R03 (AAA), 2pcs |
External Dimensions/Weight | W54×H110×D20.5mm, approx. 95g (including batteries) |
MR9504 Data Reader
- The card logger can display and print out data in trend and table formats by connecting to PC through data reader.
- Stored data can be managed file base, so data processing can be done easily by commercial spreadsheet software.
- The card logger and the data reader is connected without any physical contact through reliable infrared radiation communication.
Data Reader Specifications
(with analysis software)
Model | MR9504 |
---|---|
PC | DOS/V |
Connector Type | US series A connector |
With cable 1.4m | |
System Requirement | Windows XP sp3 or later Windows Vista (32bit) sp2 or later Windows 7 (32bit /64bit) Windows 8/8.1 (32bit/64bit) Windows 10 (32bit/64bit) |
Memory capacity: 256MB or more | |
Communication Method | Between the card logger and the data reader: Noncontact infrared radiation type |
Between the data reader and PC: USB ver2.0 conformity | |
Weight | Approx. 180g |
Contents of Analysis Software
Settings | Clock, start time, measurement interval, card NO., test NO. | ||
---|---|---|---|
Reference temperature Integration direction Logging method | Logging method Alarm temperature Keylock | Reference temperature Integration direction Alarm temperature | |
Recalling | Setting parameters, stored data | ||
Calculation | Highest, lowest, and average temperatures | ||
Calculation | ― | Calculation | |
Display | Numeric value table of all data (temperature data by year, month, day, hour and minute), trend graph | ||
Data Storage | Dedicated file format of this software, text format (csv, txt) |
*Windows is a registered trademark of Microsoft Corporation in the USA and other countries.
*MS-DOS is a registered trademark of Microsoft Corporation in the USA and other countries.