IR-CA Series Radiation Thermometers
IR-CA series is self-contained type, stationary radiation thermometer equipped with digital temperature display, parameter setting function, and measured value external output. There are general types such as low temperature type, medium to high temperature type, high-functioning type and ultra- wide type, and types for various applications, 17 types are available for non-contact temperature measurement of many fields.
Models
Main Unit
Types | Measuring Range | Models | |
---|---|---|---|
Low Temperature/Long-Wavelengtd | -50 to 100°C, 20 to 1000°C | IR-CAB | |
Low Temperature/Short-Wavelengtd | 30 to 200°C, 100 to 500°C | IR-CAE | |
Ultra-Wide | 20 to 3000°C | IR-CAW | |
By Applications | Polyester Film | 0 to 300°C | IR-CAN |
Polyetdylene Film | 30 to 300°C | IR-CAM | |
Furnace Inside | 350 to 100°C, 450 to 1300°C 500 to 1500°C |
IR-CAR | |
Silicon | 400 to 800°C, 500 to 1000°C 600 to 1200°C |
IR-CAT | |
Gallium Arsenide | 400 to 800°C, 500 to 1000°C | IR-CAU | |
Food Industry AF | 60 to 100°C | IR-CAF | |
HMD (Hot Metal Detector) | 100 to 550°C equivalent brightness temperature | IR-CADAC01 |
Setting Display Unit
Models | Specifications |
---|---|
IR-GZA0NN | Standard |
IR-GZA1NN | With emissivity remote |
IR-GZA2NN | With reflection correction function |
Specifications
For Low Temperature
Types | Low temperature/Long-wavelength | Low temperature/Short-wavelength | ||
---|---|---|---|---|
Models | IR-CAB | IR-CAE | ||
Measuring Wavelength | 8 to 13μm | 4μm | ||
Detecting Element | Pyroelectric element | PbSe | ||
Accuracy Rating | ±0.8°C | Below 200°C…±2°C 200°C or more…±1% of measured value |
±2°C | ±3°C |
Response Time | 2s | 0.2s | 0.02s | |
Emissivity Correction | Emissivity Set Value…1.999 to 0.050 | |||
Optical System | Fixed focus lens type | Focusable type | ||
Measuring Diameter | φ37/1000mm, φ20/1000mm, φ15/400mm, φ10/500mm, φ8/200mm, φ4/200m, φ40/2000mm, φ2/100mm |
φ37/1000mm φ15/400mm φ8/200mm |
Measuring distance/Distance factor Distance factor 200 |
|
Focusing System for Spot | Laser targeting, No viewfinder | Finder focusing | ||
Lens Aperture | φ15mm | φ20mm | ||
Analog Output | 4 to 20mA DC Isolated output | |||
Power Supply | 24V DC (Allowable voltage fluctuation range…22 to 28V) | |||
Power Consumption | Max. 5VA | Max. 10VA | ||
CE Marking | EMC directive EN61326+A1 Emission class A Immunity Annex A Conform only at connecting by connector |
For High Temperature
Types | Ultra-Wide |
---|---|
Models | IR-CAW |
Measuring Wavelength | 8 to 13/1.55/0.9μm |
Detecting Element | TP/InGaAs/Si |
Accuracy Rating | Below 1000°C…±5°C 1000°C or more, below 1500°C…±0.5% of measured value 1500°C or more, below 2000°C…±1% of measured value 2000°C or more…±2% of measured value (At ε≒1.0, reference operating condition ) |
Response Time | 0.1s |
Emissivity Correction | Emissivity set value…1.999 to 0.050 |
Optical System | Fixed focus lens type |
Distance Factors | |
Measuring Distance | 0.5m to ∞ |
Measuring Diameter | φ29/1000mm |
Focusing System for Spot | Finder focusing |
Lens Aperture | φ30mm |
Analog Output | 4 to 20mA DC Isolated output |
Power Supply | 24V DC (Allowable voltage fluctuation range…22 to 28V) |
Power Consumption | Max. 3VA |
CE Marking | EMC directive EN61326+A1 Emission class A Immunity Annex A Conform only at connecting by connector |
Various Applications
- IR-CAN for polyester film
Highly accurate measurement is available by using polyester polymer peculiar absorption band. - IR-CAM for polyethylene film
Highly accurate measurement is available without affected by film thickness and coloring agent by using absorption band for hydrocarbon contained in polyethylene polymer. - IR-CAR for inside furnace
The thermometer is suitable for measuring the temperature of an object inside a furnace and an object directly heated by a burner. Highly accurate measurement is available without affected by combustion gas. - IR-CAT for silicon
Temperature of a wafer can be measured accurately without being affected by the heater on the back of the wafer by using opaque wavelength band of infrared light to silicon (measuring wavelength which is absorbed by silicon). - IR-CAU for gallium arsenide
Temperature of a wafer can be measured accurately by using opaque wavelength band of infrared light to gallium arsenide (measuring wavelength which is absorbed by gallium arsenide). - IR-CAF for Food Industry AF
The thermometer can measure liquid temperature right after fruits or vegetable beverage filled a can/bottle by high-speed and highly accurately. - HMD (Hot metal Detector) IR-CADAC01
It is an exclusive device for judging the existence of metal heat mass such as aluminum. When a hot mass above the detection level is detected, the output is turned ON.
Setting Display Unit IR-GZA
It displays measured value, provides parameter setting and supplies power etc. by connecting to IR-SA or IR-CA (with RS-485 option).
Display | Temperature, status display |
---|---|
Analog Output | Output 1 4 to 20mADC (IR-GZ output, load resistance less than 600Ω) |
Event Output | Select 2 points within “High temperature alarm”,
“High-high temperature alarm”, “Low temperature alarm”, “Low-low temperature alarm” and “self diagnostic function” Relay a-contact: High, High-high, Low, Low-low Relay b-contact: Self diagnostic function Contact capacity 240V AC 1.5A 30V DC 1.5A |
Power Supply | 100 to 240V AC, 50/60Hz |
Power Consumption | 100VAC Max. 28VA, 240VAC Max. 36VA |
Working Temperature Range | -10 to 50°C |
Working Humidity Range | 20 to 90%RH (No condensation) |
Weight | Approx. 0.5kg |
*For other specifications, refer to the catalog.