IR-CZ Series Radiation Thermometers
IR-CZ series is self-contained type, stationary radiation thermometer equipped with digital temperature display, and measured value external output.
- Highly accurate measurement at high temperature range is achieved by tracing from a metal-carbon eutectic point.
- It is capable of measurement from 500°C as a single color type.
- Stability at high temperature continuous measurement and resolution from low temperature range are improved.
Models
Types | Measuring range | Models |
---|---|---|
Low to Medium Temperature | 150~1400℃ | IR-CZP2N |
Medium Temperature | 300~1600℃ | IR-CZI2N |
High Temperature | 600~3000℃ | IR-CZS2N |
High Accuracy High Temperature | 900~3500℃ | IR-CZH7N |
High Function | 400~2000℃ | IR-CZQ2N |
Specifications
SINGLE COLOR MODELS
Types | Low Temperature | Low to Medium Temperature | High Temperature |
---|---|---|---|
Models | IR-CZP | IR-CZI | IR-CZS |
Measuring wavelength | 2μm | 1.55μm | 0.9μm |
Element | InGaAs | InGaAs | Si |
Accuracy rating | Lower than 500℃…±3℃ 500℃ to 1000℃…±5℃ Higher than 1000℃…±0.5% of measurea value |
Lower than …±5℃ 1000℃ to 1500℃…±0.5% of measured value Higher than 1500℃…±0.6% of measured value |
Lower than 1000℃…±5℃ 1000℃ to 1500℃…±0.5% of measurea value 1500℃ to 2500℃…±0.6% of measurea value Higher than 2500℃…±1% of measurea value |
Response time | 3ms | ||
Distance factor | 50、200、300 | ||
Sighting | Direct view finder or Laser targeting (option) | ||
Analog output | 4 to 20mA DC isolate output, load resistance 750Ω or lower | ||
Communication interface | RS-485 Transmission of measured data, Transmission & Receiving or vaious parameters | ||
Power supply | 24V DC(Allowable voltage fluctuations range: 22 to 28V) | ||
Power consumption | Max 3.3VA | ||
Connections | Terminal or connector |
TWO COLOR MODELS
Types | Hign Temperature High Accuracy Type | High Vunction Type |
---|---|---|
Models | IR-CZH | IR-CZQ |
Measuring wavelength | 0.9/1.55μm | 0.9/1.35/1.55μm |
Element | Si/InGaAs | Si/InGaAs/InGaAs |
Accuracy rating | 900 to 3500℃ 0 to 2000℃ (200 or 300 with field diasphragm φ10mm) |
350 to 2000℃(50) 400 to 2000℃(200) 450 to 2000℃ 500 to 2000℃(200 with field disaphragm φ10mm) 550 to 2000℃(300 with field disaphragm φ10mm) |
Response time | 2 to 15ms | |
Distance factor | 200,300 | 50,200,300 |
Sighting | Direct view finder or Laser targeting (option) | |
Analog output | 4 to 20mA DC isolate output, load resistance 750Ω or lower | |
Communication interface | RS-485 Transmission of measured data, Transmission & Receiving or vaious parameters | |
Power supply | 24V DC(Allowable voltage fluctuations range: 22 to 28V) | |
Power consumption | Max 2.4VA | |
Connections | Terminal or connector |