Add finely crushed ice and water (preferably distilled water) to a thermos or other container with good heat retention.
After confirming that the ice is not melting (stable state), measure the surface temperature of the ice water.
(Please be careful to avoid any lack of view.)
At this time, it is necessary to set emissivity of radiation thermometer to 1.0.
In this state, the surface temperature of the ice water will be approximately 0°C.
Checking indicated value on radiation thermometer allows for a simple scale check.
(However, please ensure you have verified the accuracy of radiation thermometer before proceeding.)
(You need to put crushed ice into a container, dig grooves into it, and increase the apparent emissivity.)