1. Comparison method with contact thermometers
First, measure the surface temperature of the substance using a contact thermometer such as a thermocouple or resistance thermometer. Next, measure the surface temperature of the substance using radiation thermometer with emissivity set to 1.0. Gradually change emissivity of the radiation thermometer radiation thermometer are the same. emissivity at which both thermometers' values are the same is emissivity of the substance being measured.
2. How to use blackbody tape
Attach blackbody tape (or paint) with a known emissivity to the material to be measured. Set emissivity of radiation thermometer to emissivity of blackbody tape (e.g., 0.94) and measure the area where blackbody tape is attached.
Next, we measure the areas where blackbody tape is not attached.
emissivity is adjusted so that the measured value matches the value where blackbody tape is attached.
If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.
You must be a login to download.